Assessment of the Accuracy of a Virtual Multi-Channel Temperature Measuring Instrument

Authors

  • Romuald Maśnicki Gdynia Maritime University, 81-87 Morska St., 81-225 Gdynia, Poland, Faculty of Electrical Engineering, Department of Marine Electrical Power Engineering
  • Beata Pałczyńska Gdańsk University of Technology, 11/12 Gabriela Narutowicza St., 80-233 Gdańsk, Poland, Faculty of Electrical and Control Engineering, Department of Metrology and Information Systems

DOI:

https://doi.org/10.26408/127.05

Keywords:

accuracy evaluation, uncertainty, temperature measurement, multi-channel instrument, NTC thermistor

Abstract

The multi-channel temperature measurement system developed works with NTC thermistors. The article presents the results of theoretical and empirical evaluation of accuracy obtained in measurement channels. The basis for the theoretical assessment is the mathematical model for each of the measurement channels and the characteristics of the system elements included in the circuits of the measurement channel. Two different methods were used to theoretically estimate the accuracy: the Gauss Formula and the Monte Carlo method. The empirical assessment of accuracy is supported by measurement data collected during the calibration procedures for the device developed and the results of verification measurements performed on a working device. The exemplary results presented of the accuracy assessment, obtained in the selected channel, are representative of the properties of all other measurement channels. The results of both methods of assessing accuracy in measurement channels are very similar, which confirms the assumptions regarding the implementation of the instrument developed and its good metrological properties.

References

JCGM 100: Evaluation of Measurement Data — Guide to the Expression of Uncertainty in Measurement, 2008), https://www.bipm.org/documents/20126/2071204/JCGM_100_2008_E.pdf/cb0ef43f-baa5-11cf-3f85-4dcd86f77bd6?elqTrackId=2AF2E4569E5788A3689CDE7ED7E569DF&elqaid=1568&elqat=2 (14 May 2023).

JCGM 101: Evaluation of Measurement Data – Supplement 1 to GUM. Propagation of Distributions Using a Monte Carlo Method, 2008, https://www.bipm.org/documents/20126/2071204/JCGM_101_2008_E.pdf (14 May 2023).

JCGM 102: Evaluation of Measurement Data – Supplement 2 to the GUM. Extension to Any Number of Output Quantities, 2011, https://www.bipm.org/documents/20126/2071204/JCGM_102_2011_E.pdf/6a3281aa-1397-d703-d7a1-a8d58c9bf2a5?version=1.6&t=1641293321173&down-load=true (Accessed: 14 May 2023).

Liu, K.Q., Pan, X.T., Cai, J.W., 2012, Analysis of Multi-Channel Virtual Temperature Measurement System Hardware Design and Its Uncertainty, Applied Mechanics and Materials, vol. 220–223, pp. 1423–1426.

Masnicki, R., Mindykowski, J., Palczynska, B., 2022, Experiment-Based Study of Heat Dissipation from the Power Cable in a Casing Pipe, Energies, vol. 15(13).

Masnicki, R., Swisulski, D., 2023, Multi-Channel Virtual Instrument for Measuring Temperature – A Case Study, Electronics, vol. 12(10).

NI PXIe-6341, 2023, Specifications, https://www.ni.com/docs/en-US/ (14 May 2023).

NI PXIe-1082, 2023, Specifications, https://www.ni.com/docs/en-US/bundle/pxie-1082-specs/page/specs.html# (14 May 2023).

NI PXIe-8135, 2023, User Manual. National Instruments, https://www.ni.com/docs/en-US/bundle/pxie-8135rt-specs/page/specs.html (14 May 2023).

NIST Uncertainty Machine, 2023, https://uncertainty.nist.gov/ (14 May 2023).

NTC Thermistor Accuracy Temperature Sensor Thermistor NTC 10K 1% 3950 Waterproof Probe, 2017, OKYSTAR, https://www.okystar.com/product-item/ntc-thermistor-ntc-10k-oky3065-5/ (14 May 2023).

Pawłowski, E., Szlachta, A., Otomański, P., 2023, The Influence of Noise Level on the Value of Uncertainty in a Measurement System Containing an Analog-to-Digital Converter, Energies, vol. 16(3).

Samoilenko, O., 2021, The Calibration and Uncertainty Evaluation by Elementary Measurement Models, Measurements infrastructure, vol. 1, pp. 1–10.

Sârbu, G.C., 2018, Evaluation of the Measurement Uncertainty in Thermoresistances Cali­bration, International Conference and Exposition on Electrical and Power Engineering (EPE) [Preprint].

SCB-68 and SCB-68A, 2023, DAQ Multifunction I/O Accessory Guide, https://www.ni.com/en-us/support/documentation/supplemental/17/scb-68-and-scb-68a---daq-multifunction-i-o-accessory-guide.html (14 May 2023).

Steinhart, J.S., Hart, S.R., 1968, Calibration Curves for Thermistors, Deep Sea Research and Oceanographic Abstracts, vol. 15(4), pp. 497–503.

Thermometer DT-34, 2023, High Accuracy Resistance Food Industrial Min Max Probe Digital Thermometer DT-34. Display Temperature Meter with 0.1°C Accuracy for Agriculture, Labs and Food.(-100°C to 270°C), https://termoprodukt.co.uk/industrial-thermometer-dt-34 (14 May 2023).

Downloads

Published

2023-09-30

How to Cite

Maśnicki, R., & Pałczyńska, B. (2023). Assessment of the Accuracy of a Virtual Multi-Channel Temperature Measuring Instrument. Scientific Journal of Gdynia Maritime University, (127), 57–70. https://doi.org/10.26408/127.05